LIQUIDROME
Liquids and degassing samples equipment
Methods
Remote access
not possible
Station data | |
---|---|
Temperature range | 283 - 373 K |
Pressure range | -- |
Detector | time-of-flight (TOF) electron energy analyzer, GaAs photodiode |
Manipulators | motorized XYZ sample manipulator with micrometer precision |
Sample holder compatibility | -- |
Additional equipment | |
Chamber pressure | 10^-5 mbar base pressure in main chamber and 10^-8 mbar in TOF analyzer when liquidjet running in main chamber |
Liquidrom is a vacuum system dedicated to x-ray photoelectron spectroscopy (XPS) and x-ray absorption spectroscopy (XAS) for liquid samples. The system is equipped with time-of-flight (TOF) electron energy analyzer and GaAs photodiode as electron and photon detectors respectively. Intense laser pulses, synchronized with the synchrotron x-ray pulses (only in single bunch mode), are often employed in Liquidrom system to excite (pump) samples by laser and successively (within nanoseconds) measure (probe) samples by x-ray in order to study the dynamic properties of various liquid phase molecules in strong laser field (Pump-Probe experiment)
For more details and current status of the experimental station contact the station scientist.