• Hwang, J-G.; Yang, T.-K.; Forck, P.; Noh, S.Y.; Hahn, G.; Choi, M.: Development of high resolution linear-cut beam position monitor for heavyion synchrotron of KHIMA project. Nuclear Instruments & Methods in Physics Research A 852 (2017), p. 33-39

10.1016/j.nima.2017.01.073
Open Access Version

Abstract:
A beam position monitor with high precision and resolution is required to control the beam trajectory for matching to the injection orbit and acceleration in a heavy-ion synchrotron. It will be also used for measuring the beta function, tune, and chromaticity. Since the bunch length at heavy ion synchrotron is relatively long, a few meters, a boxlike device with plates of typically 20 cm length is used to enhance the signal strength and to get a precise linear dependence with respect to the beam displacement. Especially, the linear-cut beam position monitor is adopted to satisfy the position resolution of 100 μm and accuracy of 200 μm for a nominal beam intensity in the KHIMA synchrotron of ∼7 ×108 particles for the carbon beams and ∼2 ×1010 for the proton beams. In this paper, we show the electromagnetic design of the electrode and surroundings to satisfy the resolution of 100 μm, the criteria for mechanical aspect to satisfy the position accuracy of 200 μm, the measurement results by using wire test-bench, design and measurement of a high input impedance pre-amplifier, and the beam-test results with long (∼1.6 μs) electron beam in Pohang accelerator laboratory (PAL).