TOF-PES

Magnetic Bottle Time of Flight - PES Set-up

Methods

XPS

Remote access

not possible

Beamline data
Energy range 85 - 1600 eV
Energy resolution Energy and slit dependent, resolving power 4300 at 400 eV and slit 20 µm
Flux see the flux curve below
Polarisation horizontal
Focus size (hor. x vert.) slit dependent, see the "focal size" figure below
Phone +49 30 8062 14835
More details U49-2_PGM-1
Station data
Temperature range 0 °C - 95 °C
Pressure range working atmosphere: 10-3 - 10-7 mbar
Detector Magnetic-bottle time-of-flight electron analyzer
Manipulators for liquid microjets: temperature controlled (0-95 °C), xyz-axis adjustable, micrometer-precision
Sample holder compatibility flexible on an ISO-K 100 flange
Additional equipment